National Instruments NI 4070/4072 6 1/2-Digit FlexDMM Graphics Tablet User Manual


 
NI 4070/4072 Calibration Procedure 28 ni.com
35. Call niDMM_ConfigureMultiPoint with the following parameters:
Trigger Count =
1
Sample Count = 10
Sample Trigger = NIDMM_VAL_IMMEDIATE
Sample Interval = –1
36. Call niDMM_ReadMultiPoint with the following parameters:
Maximum Time =
NIDMM_VAL_TIME_LIMIT_AUTO
Array Size = 10
Average the results by summing the returned reading array of the
function and dividing by the returned actual number of points.
Store the result as the 100 Ω range offset.
37. Output 100 MΩ on the calibrator without external sense or 2-wire
compensation.
38. Call
niDMM_ConfigureMeasurement with the following
parameters:
Function =
NIDMM_VAL_2_WIRE_RES
Range = 100e6
Resolution = 100
39. Call niDMM_Read. Subtract the previously stored 100 MΩ range offset
from this measurement. Verify that the result falls between the
tolerances listed in Table 18.
40. Output 10 MΩ on the calibrator without external sense or 2-wire
compensation.
41. Call
niDMM_ConfigureMeasurement with the following
parameters:
Function =
NIDMM_VAL_2_WIRE_RES
Range = 10e6
Resolution = 10
42. Call
niDMM_Read. Subtract the previously stored 10 MΩ range offset
from this measurement. Verify that the result falls between the
tolerances listed in Table 18.
43. Output 1 MΩ on the calibrator without external sense or 2-wire
compensation.
44. Call
niDMM_ConfigureMeasurement with the following
parameters:
Function =
NIDMM_VAL_2_WIRE_RES
Range = 1e6
Resolution = 1