© National Instruments Corporation 47 NI 4070/4072 Calibration Procedure
58. Call niDMM_CalAdjustOffset with the following parameters:
• Mode =
NIDMM_VAL_2_WIRE_RES
• Range = 1e6
• Input Resistance = NIDMM_VAL_RESISTANCE_NA
59. Call niDMM_CalAdjustOffset with the following parameters:
• Mode =
NIDMM_VAL_2_WIRE_RES
• Range = 100e3
• Input Resistance = NIDMM_VAL_RESISTANCE_NA
60. Call niDMM_CalAdjustOffset with the following parameters:
• Mode =
NIDMM_VAL_2_WIRE_RES
• Range = 10e3
• Input Resistance = NIDMM_VAL_RESISTANCE_NA
61. Call niDMM_CalAdjustOffset with the following parameters:
• Mode =
NIDMM_VAL_2_WIRE_RES
• Range = 1e3
• Input Resistance = NIDMM_VAL_RESISTANCE_NA
62. Call niDMM_CalAdjustOffset with the following parameters:
• Mode =
NIDMM_VAL_2_WIRE_RES
• Range = 100
• Input Resistance = NIDMM_VAL_RESISTANCE_NA
63. Call niDMM_CalAdjustMisc with Type set to
NIDMM_EXTCAL_MISCCAL_SECTION.
You have completed adjusting the DC voltage and resistance modes of the
NI 4070/4072. Select one of the following options:
• If you are performing additional adjustments, refer to the following
sections, as applicable:
– Adjusting AC Voltage (AC- and DC-Coupled) Modes
– Adjusting Current Modes
– Adjusting Capacitance and Inductance (NI 4072 Only)
Caution For the NI 4072, adjusting the capacitance and inductance is required. Skipping
this step causes an incorrect adjustment of the device.