© National Instruments Corporation 47 NI 4070/4072 Calibration Procedure
58. Call niDMM_CalAdjustOffset with the following parameters:
Mode =
NIDMM_VAL_2_WIRE_RES
Range = 1e6
Input Resistance = NIDMM_VAL_RESISTANCE_NA
59. Call niDMM_CalAdjustOffset with the following parameters:
Mode =
NIDMM_VAL_2_WIRE_RES
Range = 100e3
Input Resistance = NIDMM_VAL_RESISTANCE_NA
60. Call niDMM_CalAdjustOffset with the following parameters:
Mode =
NIDMM_VAL_2_WIRE_RES
Range = 10e3
Input Resistance = NIDMM_VAL_RESISTANCE_NA
61. Call niDMM_CalAdjustOffset with the following parameters:
Mode =
NIDMM_VAL_2_WIRE_RES
Range = 1e3
Input Resistance = NIDMM_VAL_RESISTANCE_NA
62. Call niDMM_CalAdjustOffset with the following parameters:
Mode =
NIDMM_VAL_2_WIRE_RES
Range = 100
Input Resistance = NIDMM_VAL_RESISTANCE_NA
63. Call niDMM_CalAdjustMisc with Type set to
NIDMM_EXTCAL_MISCCAL_SECTION.
You have completed adjusting the DC voltage and resistance modes of the
NI 4070/4072. Select one of the following options:
If you are performing additional adjustments, refer to the following
sections, as applicable:
Adjusting AC Voltage (AC- and DC-Coupled) Modes
Adjusting Current Modes
Adjusting Capacitance and Inductance (NI 4072 Only)
Caution For the NI 4072, adjusting the capacitance and inductance is required. Skipping
this step causes an incorrect adjustment of the device.