National Instruments NI 4070/4072 6 1/2-Digit FlexDMM Graphics Tablet User Manual


 
NI 4070/4072 Calibration Procedure 34 ni.com
14. Call niDMM_ConfigureMeasurement with the following
parameters:
Function =
NIDMM_VAL_AC_CURRENT
Range = 1
Resolution = 1e–6
15. Call niDMM_Read. Verify that this measurement falls between the
limits listed in Table 20.
16. Output 1 A at 1 kHz on the calibrator.
17. Call
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 20.
You have completed verifying the AC current of the NI 4070/4072. Select
one of the following options:
If you want to continue verifying other modes, go to the Verifying
Frequency section.
If you do not want to verify other modes and you are performing a
post-adjustment verification, go to the Completing the Adjustment
Procedures section.
If you do not want to verify any additional modes and you are
performing a pre-adjustment verification, call
niDMM_close to close
the session.
Verifying Frequency
Notes The frequency of the NI 4070/4072 is not user adjustable. If this verification
procedure indicates that the frequency is out of specification, return the NI 4070/4072
to NI for repair.
This verification procedure is optional and requires additional test
equipment. If you do not want to verify frequency, select one of the
following options:
If you are calibrating an NI 4072 and want to continue verifying other
modes, go to the Verifying Capacitance and Inductance (NI 4072
Only) section.
If you do not want to verify other modes and are performing a
post-adjustment verification, go to the Completing the Adjustment
Procedures section.
If you do not want to verify any additional modes and you are
performing a pre-adjustment verification, call
niDMM_close to close
the session.