National Instruments NI 4070/4072 6 1/2-Digit FlexDMM Graphics Tablet User Manual


 
© National Instruments Corporation 45 NI 4070/4072 Calibration Procedure
36. Call niDMM_CalAdjustGain with the following parameters:
Mode =
NIDMM_VAL_2_WIRE_RES
Range = 100e6
Input Resistance = NIDMM_VAL_RESISTANCE_NA
Expected Value = the display on the calibrator for 0 Ω
37. Call
niDMM_CalAdjustOffset with the following parameters:
Mode =
NIDMM_VAL_2_WIRE_RES
Range = 100e6
Input Resistance = NIDMM_VAL_RESISTANCE_NA
38. Output 100 kΩ on the calibrator with external sense turned on
but without 2-wire compensation.
39. Call
niDMM_CalAdjustGain with the following parameters:
Mode =
NIDMM_VAL_4_WIRE_RES
Range = 100e3
Input Resistance = NIDMM_VAL_RESISTANCE_NA
Expected Value = the display on the calibrator for 100 kΩ
40. Output 0 Ω on the calibrator with external sense turned on but without
2-wire compensation.
41. Call
niDMM_CalAdjustGain with the following parameters:
Mode =
NIDMM_VAL_4_WIRE_RES
Range = 100e3
Input Resistance = NIDMM_VAL_RESISTANCE_NA
Expected Value = the display on the calibrator for 0 Ω
42. Output 10 kΩ on the calibrator with external sense turned on
but without 2-wire compensation.
43. Call
niDMM_CalAdjustGain with the following parameters:
Mode =
NIDMM_VAL_4_WIRE_RES
Range = 10e3
Input Resistance = NIDMM_VAL_RESISTANCE_NA
Expected Value = the display on the calibrator for 10 kΩ
44. Output 0 Ω on the calibrator with external sense turned on but without
2-wire compensation.
45. Call
niDMM_CalAdjustOffset with the following parameters:
Mode =
NIDMM_VAL_4_WIRE_RES
Range = 100e3
Input Resistance = NIDMM_VAL_RESISTANCE_NA