SCPI Command Reference
SCPI Command Summary
Release 2.1 107
Table B.4Commands for Current Share
Table B.5Commands for Calibration
Table B.6Command to Clear all Protection Mechanisms
Set High Voltage Limit
[[:]SOURce][<channel>]:VOLTage:LIMit:HIGH
{<voltage>|MAXimum|MINimum}
Set upper limit of voltage setpoint range (soft
limits)
Yes
Set Low Voltage Limit
[[:]SOURce][<channel>]:VOLTage:LIMit:LOW
{<voltage>|MAXimum|MINimum}
Set lower limit of voltage setpoint range (soft
limits)
Yes
Function
SCPI Command
Description Query
Set Current Sharing Mode
[[:]SOURce][<channel>]:COMBine:CSHare:MODE
{NONE|MASTer|SLAVe}
Select current share mode Yes
Read Summed Current
[:]MEASure[:SCALar]:CURRent[:DC]? SUM
Read total current output of all current
sharing supplies
N/A
Function
SCPI Command
Description Query
Restore Factory
ion
[:]
CALibration[<channel>]:RESTore
Restonres the calibration to the constants set
at the factory
N/A
Change Calibration
Password
[:]
CALibration[<channel>][:SECure]:CODE <codeword>
Changes the calibration security code. No
Set Calibration State
[:]
CALibration[<channel>][:SECure]:STATe
<on-off-state>,<codeword>
Change calibration state (mode) Yes
Set Supply Output Current
Level
[:]
CALibration[<channel>]:OUTPut:CURRent:LEVel
{MINimum|MAXimum}
Set output current calibration level No
Enter Output Current Data
[:]
CALibration[<channel>]:OUTPut:CURRent[:DATA]
<current>
Set output current calibration data No
Set Supply Output Voltage
Level
[:]
CALibration[<channel>]:OUTPut:VOLTage:LEVel
{MINimum|MAXimum}
Set voltage output calibration level No
Enter Output Voltage Data
[:]CALibration[<channel>]:OUTPut:VOLTage[:DATA]
<voltage>
Set voltage output calibration data No
Function
SCPI Command
Description Query
Clear Output Protection
[:]OUTPut[<channel>]:PROTection:CLEar
Clears the protection mechanism. N/A