Xantrex Technology GPIB-M-XT Power Supply User Manual


 
Operation
Power Supply Operation
Release 2.1 57
Power On
Output State
(XFR/XHR only) You may also change the output state whether the output is enabled
or disabled at power on.
To change the power on at output state:
OUTP:PON:STAT [ON|OFF|1|0]
Whether the unit powers on with output enabled is a function of both the state at
power off and this setting.
Reset Resetting the unit puts certain features to a known state. These states are listed in
Table 3.2.
To reset the unit:
*RST or SYST:RES
Table 3.2 Features Affected by Reset (*RST) Command
Feature Reset State (*RST)
Voltage setpoint 0.0V
Current setpoint 0.0A
Low voltage setpoint limit 0.0V
High voltage setpoint limit 101% of voltage rating
Low current setpoint limit 0.0A
High current setpoint limit 101% of current rating
Over voltage protection Disabled (0.0V)
Under voltage protection Disabled (0.0V) and not shutdown when tripped
Over current protection Disabled (0.0A) and not shutdown when tripped
Under current protection Disabled (0.0A) and not shutdown when tripped
Fold shutdown protection None and delay 0.5s
AC off shutdown recovery Auto recover
OTP shutdown recovery Latched
Triggered voltage setpoint Disabled (Default)
Triggered current setpoint Disabled (Default)
Autosequence mode Exit
Calibration mode False
Output Disabled