RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL
© 2007 DH Instruments, a Fluke Company Page 70
Table 9. Reset – Sets
RESET RESULT SEE SECTION
[UNIT] Pressure unit of measure to first of the six available (psi) 3.3.2
[MODE] Measurement mode to the native mode of the Q-RPT (absolute) 3.3.3
[RANGE] Hi Q-RPT default range (10000 psi) 3.3.1
[HEAD] 0 cm height and Nitrogen medium 3.3.7
Stability Limit 0.005 % of Hi Q-RPT default range 3.4.3
Upper Limit Hi Q-RPT default range default value 3.4.4
Resolution 0.001 % FS of Hi Q-RPT default range 3.4.2
AutoZ AutoZ ON for both Q-RPTs and gauge and asbsolute
measurement modes. Z
offset
value not affected
3.5.1
AutoZ P
atm,0
set to 101.325 kPa a 3.5.1, 3.2.2
Leak Check 15 second run time. Clear logged results 3.3.5
Screen Saver 10 minutes to activation 3.5.5.1
Key Sounds Medium tone valid key press sound 3.5.5.2
Lo Vnt Automatic 3.5.7.4
ReadRate Automatic 3.5.7.2
3.5.9.2 <2 UNITS>
PURPOSE
Sets the pressure units available under the UNIT function to default selections
depending (see Section 3.3.2).
Sets the user defined unit to 1.000/Pa (see Section 3.5.6).
Sets the reference temperature for inWa unit to 20°C.
3.5.9.3 <3ATEST>
PURPOSE
Resets Auto Tests to default characteristics (see Section 3.3.10).
Clears File AutoTest definition files (see Section 3.4.5.2).
Clears AutoTest results data logs (see Section 3.4.5.1).