4. REMOTE OPERATION
Page 77 © 2007 DH Instruments, a Fluke Company
4.4 COMMANDS
4.4.1 PROGRAMMING MESSAGES
Some commands accept an optional suffix. This suffix can be used to specify the explicit Q-RPT
to address. In most cases, each Q-RPT in the RPM4 operates as an independent measurement
device. If the suffix is not specified, then the “Active” Q-RPT is assumed.
The “RANGE” or “ARANGE” command can be used to select the “Active” Q-RPT. See the
program message descriptions for suffix use with a specific command.
Table 14. Program message list
COMMAND
n’ indicates an
optional suffix
DESCRIPTION
# Send a command string out of the RPM4 COM2 port.
ABORT Stop pending operations
ARANGEn Read or set a new AutoRange to use.
ATM Read the current atmospheric pressure (on-board barometer).
AUTOZEROn Read or set the status of the AutoZ automatic zeroing function.
AUTOZEROn=RUN Run AutoZero.
CALAMB Read or set the on-board barometer calibration
COM1 Read or set the configuration of the COM1 port.
COM2 Read or set the configuration of the COM2 port.
CONTn Enable continuous measurements
DATE Read or set the current date.
ERR Read the last error message.
GPIB Read or set the GPIB interface address.
HEAD Read or set the fluid head settings.
ID Read or set the RPM4 alphanumeric asset ID tag.
L2 Selects “classic” program message format
L3 Selects “enhanced” program message format
LOCAL Return control to the RPM4 front panel.
MEM Read the power-up memory test status.
MMODEn Read or change the active measurement mode.
MSGFMT Read or set the type of program message format to use.
PCALn Read or set the user Lo Q-RPT calibration information.
PCAL:XX Read or set the user Lo Q-RPT calibration information (old command).
PRn Read the next RPM4 pressure.
PRRn Read the next RPM4 pressure, rate, and ATM.
QPRRn Read the last RPM4 pressure, rate and ATM.
RANGE Select a Q-RPT to be “active” in it’s default full scale range
RATEn Read the next available rate of change of pressure.
READRATEn Read or set the Q-RPT measurement read rate and mode
READYCKn Read or set a flag that is cleared by a Not Ready condition.
REMOTE Enable remote local lockout operation.
RESn Read or set the pressure display resolution for the current transducer and range.
RESET Reset the RPM4 to default user parameters.
RPTn Read the available Q-RPT data.
SCRSAV Read or set the front panel screen saver period.
SN Read the serial number of the RPM4.
SRn Read the next available pressure status (Ready/Not Ready).
SSn Read or set the stability required for a Ready condition.
SS%n Read or set the stability required for a Ready condition (% span/ s).
TIME Read or set the current time of day.
UCOEFn Convert a pressure in Pascal to pressure in the current units.
UDU Read or set the user defined pressure unit.
ULn Read or set the upper limit for the current range.
UNITn Read or set the pressure unit of measure for the current range.
VER Read the RPM4 software version.
ZOFFSETn Read or set the AutoZero Z
offset
for the specified Q-RPT.
ZOFFSET:XX Read or set the AutoZ Z
offset
for the specified Q-RPT (old command).