Intelligent Motion Systems UDAS-1001E Computer Hardware User Manual


 
Configuration and Installation
2-4
Channels connected to a thermocouple must be connected to bias resistors for proper operation.
Use jumpers JP2 and JP3.
To select the jumpers for each channel that you plan to use for thermocouples, use TABLE 2.1,
CJC Jumper Positions shown below.
TABLE 2.1 CJC Jumper Positions
Note: When monitoring thermocouples, always use differential channels.
Differential JP1 Jumper Position JP2 Jumper Position
Channel #
1 AI 1 AI 9
2 AI 2 AI 10
3 AI 3 AI 11
4 AI 4 AI 12
5 AI 5 AI 13
6 AI 6 AI 14
7 AI 7 AI 15