Troubleshooting
2-6 March 2001
205 Must enter the following
field:_field_.
The operator left a configuration
field blank.
Enter information in the blank
configuration field.
301 The following key was not
found: _key_.
• An internal error occurred with
the software.
• One of the files was edited
manually and a key was
changed.
• Restart the application.
• Restore the key to its previous
value.
• If problem continues, call for
service.
302 The following grouper was
not found:_grouper_.
• An internal error occurred with
the software.
• One of the files was edited
manually and a grouper was
changed.
• Restart the application.
• Restore the grouper to its
previous value.
• If problem continues, call for
service.
303 The call depends on a
current grouper, but the
current grouper has not been
set.
An internal error occurred with the
software.
• Restart the application.
• If problem continues, call for
service.
501 An error occurred while
applying the data.
601 A bad count was entered.
The expected count
was:_count_.
• A file was edited manually and
an incorrect number of rows
was entered.
• An internal software error
occurred.
• Manually enter the correct
number of rows in the file.
• If problem continues, call for
service.
602 An error occurred during the
calculation process.
An internal error occurred with the
software.
• Restart the application.
• If problem continues, call for
service.
701 Patch _patch_, column
_column_, value _value_ is
beyond tolerance
_tolerance_ for aim _aim_.
A density patch was out of
tolerance.
Perform another calibration cycle.
Code Status Message Possible Cause/Subsystem Possible Solution