36
Chapter4SelfCalibration
Chapter 4 SelfCalibration
This product is equipped with a built-in calibration sensor. The conditions that have been adjusted using
ColorNavigator and the measurement device can be maintained.
As the adjustment targets or interval times for SelfCalibration are set in advance using ColorNavigator, the built-in
calibration sensor automatically executes regular calibration.
Monitor operation and setting are not necessary when ColorNavigator is used.
If ColorNavigator cannot be run due to the usage environment, SelfCalibration can be executed for an independent
monitor.
Note
• The measurement results from the built-in calibration sensor can be correlated with the measurement results from the
measurement device in use. For details, refer to the ColorNavigator User’s Manual (on the CD-ROM).
• SelfCalibration can also be executed when there are no computer signals being input.
4-1. Setting Calibration Details
●
Setting monitor date and time
Set the monitor date and time.
Note
• When ColorNavigator is started, the date and time are automatically set. For details, refer to the ColorNavigator User’s
Manual (on the CD-ROM).
Procedure
1. Choose “SelfCalibration” from the Adjustment menu, and press
.
2. Choose “Settings” from “SelfCalibration”, and press
.
3. Choose “Clock Adjustment” from “Settings”, and press
.
The Clock Adjustment menu appears.
4. Set date with or , and press .
5. Set time with
or , and press .
Attention
• If the main power supply is disconnected for an extended period of time, the clock may need resetting.