Eizo CG276 Computer Monitor User Manual


 
39
Chapter4SelfCalibration
Setting Adjustment Target
Set the calibration targets for SelfCalibration.
Procedure
1. Choose “SelfCalibration” from the Adjustment menu, and press
.
2. Choose “Settings” from “SelfCalibration”, and press
.
3. Choose “Target” from “Settings”, and press
.
4. Use
or to select the mode to change, and press .
The target setting menu appears.
5. Use or to select the items to change, and press .
Settings Adjustable Range Setting
Brightness 30cd/m
2
to200cd/m
2
Setthebrightnessofthemonitorusing
thebrightnessvalue(cd/m
2
).
WhitePoint White(x)/White(y) 0.2400-0.4500
*1
SetthemonitorWhitePointusingthe
colorcoordinateorTemperature.
Temperature 4000K-10000K
(100KUnit)
*2
ColorGamut ColorGamut Native Donotadjustthecolorgamut.
User Adjustthecolorgamut.
ForeachofRed/Green/
Blue
0.0000-1.0000 Tocalibratethemonitorcolor
reproductiongamut,setthecolor
coordinatesforred,green,andblue.
Clipping On/Off Thisfunctionallowsyoutodisplaythe
monitorwiththecolorsbasedonthe
speciedcolorgamut.
(“Tosetclipping”(page22))
Gamma 1.6-2.7
(0.1Unit)
*3
Setthemonitorgammavalues.
*1
 Whenthecolorcoordinatesareset,thetemperaturechangesto“User”.
*2
 Whenthetemperatureisset,thexandycoordinatesofthewhitepointaresetautomatically.
*3
 Thesebecome“Fixed”whencalibratingwithColorNavigator.
6. Use or for adjustment.
7. Press
to exit.
The calibration targets are set.
Attention
Only set the color gamut calibration when the target color range is precisely known.
If ColorNavigator was used to calibrate the grey balance, do not change the gamma setting values.
4-2. Executing SelfCalibration
Procedure
1. Choose “SelfCalibration” from the Adjustment menu, and press
.
2. Choose “Execute” from “SelfCalibration”, and press
.
The built-in calibration sensor is activated, and SelfCalibration is executed.
Attention
To execute SelfCalibration, it is necessary to set the calibration details. For more information, see “4-1. Setting
Calibration Details” (page 36).
After selecting “Execute”, aging may be executed before the built-in calibration sensor is executed.