Quantum Instruments Cascade Laser Starter Kit Power Supply User Manual


 
Starter Kit Instruction Manual Description 18
4.4.3 Thermo-Electric Cooler (TEC)
and PT100 connections
The pin 1 is re c ognised by a circle around it and
a thin noon ended circle starts from pin one and
turns around the pins until pin 6.
(1) Pin 1: +Peltier element
(2) Pin 2: - Peltier element
(3) Pin 3: I+ PT100
(4) Pin 4: Sens + PT100
(5) Pin 5: Sens - PT100
(6) Pin 6: I- PT100
Specifications of the Peltier connector
Vacuum case plug reference: .. . . . . .LEMO #
HGG.1b.306.CLLPV.
Cable plug reference: . . . . . . . .. . . . . LEMO #
FGG.1b.306.CLAD72Z.
Plastic protection for the c able: . . .LEMO #
GMA.1b.065.DG.
Cable:
Diameter : . . . . . .. . . . . . . .. . . . . .. 5.8 mm
Section : 6 x 0.25 mm (2 x 0.25 for each
Peltier connections)
or
Diameter : . . . . . .. . . . . . . .. . . . . .. 7.2 mm
Section : 10 x 0.25 mm (2 x 3 x 0.25 for
Peltier connections)
4
5
6
1
2
3