Starter Kit Instruction Manual Description 21
4.6 QCL pulse switching measuring unit
4.6.1 Generalities
The measuring circuit c ontained in the LDD100
provides information about various laser pulse pa-
rameters (peak voltages, supply voltag e s, duty cy-
cle, frequency). They are only estimated values,
since exact measurement of short and strong pulses
with diodes and averaging circuitry is difficult.
Keep in mind that you should always measure
the voltages on the LLH100 if you need accurate
time and voltage data (see final paragraph). How-
ever, these data are useful for monitoring and surveil-
lance purposes, and to give a rough estimation of
the current parameters.
4.6.2 Measured voltages
UHV: 1/2 of the average voltage, respective
to VHT (user supplied high voltage)
ULH: 1/2 of the average laser anode voltage,
respective to VHT
ULL: duty cycle depe ndent peak laser cathode
voltage, respective to VHT
UD: duty cycle dependent peak trans istor drain
voltage, respective to VHT
UPI: 1/2 average internally reshaped drive volt-
age, respective to ground
UPT: 1/2 average of a 37ns fixed length pulse,
respective to ground
4.6.3 Timing data
Pulse frequency calculation
UPT can be used to calculate the pulse frequency
as follows:
f = f
0
2UPTν
Plow
ν
Phigh
ν
Plow
Where f
0
is a frequency constant, v
Plow
and
v
Phigh
are the TTL pulser voltage limits.
Actual values: f
0
=
1
37
ns , v
Plow
=0.01V, v
Phigh
=5V