Quantum Instruments Cascade Laser Starter Kit Power Supply User Manual


 
Starter Kit Instruction Manual Maintenance 51
Current adjustment
[28] Set the selector (3) to mode Setting
C
and adjust the temperature value to 25
C
by means of the Set Temperature 5 tuns
portentiometer P8 (2).
[29] On the external p ower supply, rise slowly
the voltag e to 30 V and the current to 1.5 A.
The tension should stabilize at 30 V
Note: If the current is too high, stop the test
and verify the transistors Q1 to Q4 (1) and
their associated resistors.
There might be a soldering problem. In this
case, fix it and repeat the procedure from step
[10].
[30] Set the selector (3) to mode Real I.
Value displayed: 1A.
[31] Set the selector (3) to mode Real
C.
Value displayed should tend to the defined
value.
[32] Wait for a certain time and check the
Peltier temperature with a thermo meter
[33] Set the selector (3) to mode Setting +I.
[34] Increase the limit of positive current to
3.5A by means of the +I trimmer (4) located
on the front panel.
[35] Increase the limit of negative current to
5A by means of the -I trimmer (5) located on
the front panel.
[36] Change the temper ature refer ence to -25
C and check the displayed current.
1
Fig.47: TCU151 main board
TCU151
T
e
m
p
e
r
a
t
u
r
e
C
o
n
t
r
o
l
l
e
r
A
l
a
r
m
R
e
s
e
t
E
x
t
e
r
n
a
l
I
n
t
e
r
n
a
l
R
e
a
l
I
R
e
a
l
C
S
e
t
t
i
n
g
C
S
e
t
t
i
n
g
+
I
S
e
t
t
i
n
g
I
+
5
4
3
2
Fig.48: TCU151 front panel