4.4 Self-check
(1) Self-check Procedure
To run the emulator self-check, do so as explained here below. While the self-check is in progress,
LEDs will change as shown in Figure 4.5.
(1) Set the switches in the emulation pod same as the factory setting (see Figure 4.4).
(2) When the target system is connected, disconnect the target system.
(3) Within 2 seconds of activating power to the emulator, press the RESET button on the emulator
front panel to switch the emulator to maintenance mode.
(4) Check the "SAFE" LED starts flashing and then press the RESET button again.
(5) The self-check will start. If the normal result is displayed in about 1 minute, the self-check has
terminated normally.
Figure 4.4 Switch settings when running the self-check
(2) If an Error is Detected in the Self-check
If the self-check does not result normally (ERROR 1 or ERROR 2 in Figure 4.5), check the following.
• Check the connection of the emulation pod and the PC4701.
•Redownload the proper firmware.
• Check if the switches in this product is set same as the factory-setting ("Figure 3.2 Positions of
switches and their factory-settings", page 25).
IMPORTANT
Note on Self-check:
If the self-check does not result normally (excluding target system errors), the
emulation pod may be damaged. Contact your local distributor.