■ Weighing with manually
entered tare
If the patient is already on the bed and the
bed weight is known, then this can be tared by
manual input.
Enter the known bed weight (tare weight) using
the KEYPAD. The input appears in the display with
the current weight unit. Press the TARE key. The
nput is transferred into the tare memory and the
tare weight is deducted from the weight display.
The patient's weight measured in this way is dis-
played with the message "Net".
Showing the value in the tare memory
Display the tare memory by pressing the INFO key,
then press the Tare key. Press the ZEROING key to
end the display.
Changing the tare value manually later
Show the tare value by pressing the INFO key and
then the TARE key. Enter the new value into the
KEYPAD. Press the ZEROING key to cancel the input.
The existing value is retained or the new tare value
is saved by pressing the TARE key. Confirm by pres-
sing ENTER and return to the application program.
■ Weighing with stored tare values
Known bed weights can be stored as the tare value
in a power failsafe memory and called up as required
(100 different bed weights are possible).
Each memory location has a:
• Three-digit consecutive number
• Description (text, 18 characters)
• Tare value
Entering tare weights into the fixed tare
value memory
• Call up the input menu
(Setting mode > Program Settings > General
> Fixed tare values)
• Press ENTER and F key simultaneously to access
Setting mode.
• The arrow points to "Program Settings". Confirm
by pressing ENTER.
• Confirm the message "General" by pressing ENTER.
• Press TARE to browse to "Fixed tare values" and
accept by pressing ENTER.
Entering values
• If no tare memory has yet been assigned, the
first free, or otherwise the last, selected fixed
tare value memory is displayed.
• Press ORG followed by ENTER. The input field
for the description of a new tare memory is dis-
played.
• Enter the description into the KEYPAD and con-
firm by pressing ENTER.
• The arrow points to "Tare value". Press ENTER.
The input field is displayed to enter the tare value.
• Enter the tare value and confirm by pressing
ENTER.
• Go back by pressing the ZEROING key to select
the tare memory.
• If you want to assign other tare memories, press
the ORG key to select the next free memory lo-
cation and repeat the procedure as described
above.
• To return to weighing mode, press the ZEROING
key and then the ORG key.