Soehnle 7711 Postal Equipment User Manual


 
ENGLISH
Bed scale 7711
4
Safety instructions
Before putting the appliance into use,
please read with care the information given in the
Operating Instructions. They contain important in-
structions for installation, intended use and mainte-
nance of the appliance. The manufacturer shall
not be liable if the following instructions are
not observed:
In case the electrical components are used under
increased safety requirements, compliance with the
appropriate regulations is compulsory. Never perform
work on the appliance while the power is switched
on. Improper installation will render the warranty null
and void. Ensure the voltage marked on the power
supply unit matches your mains power supply. The
device is designed for use within buildings. Observe
the permissible ambient temperatures for use (Tech-
nical Information). The device meets the requirements
for electromagnetic compatibility. Do not exceed the
maximum values specified in the applicable stand-
ards. If you have any problems, contact your local
Soehnle Professional service partner.
Warning: A change of the medicine device is pro-
hibited. This device may not be changed without the
manufacturer's authorisation. If the medicine device
is changed, appropriate analysis and examination
must be undertaken to ensure a safe continued use.
Use of additional devices
Only additional appliances (printer, computer) may
be connected to the serial data port if they conform
to European Standard EN 60601-1 or a suitable dis-
connecting device is connected in series.
Applicator
The scale does not have an applicator. No part is in-
tended for direct contact with the patient. Leakage
currents were measured on the weighing devices in
compliance with classification of Type B applicators.
Operating concept
The appliance is operated by means of the keypad.
The keys are assigned with different functions de-
pending on the various modes.
Maintenance and service
The measuring inspection test was performed during
the initial calibration process in production. Further
regular measuring inspection tests (recalibration) must
be carried out by the calibration authorities respon-
sible in compliance with national regulations. Repairs
may only be carried out by a centre authorised by
Soehnle Professional using genuine spare parts.
The medicine device is equipped with a non-de-
tachable power line. The power line may not be ex-
changed by maintenance personal.
Cleaning
Before cleaning, disconnect the appliance from the
mains by removing the power plug from the wall
socket. Only clean the appliance using a moist cloth.
On no account may water be allowed to enter into
the appliance. Disinfectant may only be applied to
the membrane keyboard for the display. It is prohi-
bited to spray the appliance and the power plug.
The following disinfectants are permitted: methy-
lated spirits, Isopropanol; 2% Kohrsolin, 1% aqueous
Sokrena solution, 5% Sagrotan or 5% Gigasept.