Reference
22
80A03 TekConnect Probe Interface Module Instruction Manual
CAUTION. To pr event damage from electrostatic discharge, install 50 Ω
terminations (Tektronix part number 015-1022-xx) on the sampling-module
connectors before removing the sampling modules from an instrument or when it
is not in use. Store the sampling module in a static-free container, such as the
shipping container. Whenever you move the sampling module from one instru-
ment to another, use a static-free container to transport the sampling module
To prevent damage to the sampling module, discharge to ground any electrostat-
ic charge that may be present on the center and outer conductors of cables
before attaching the cable to the sampling module.
To prevent damage to the sampling module, do not create an ESD antenna by
leaving cables attached to the sampling-module input with the other end of the
cable open.
To prevent damage to the sampling module or instrument, never install or
remove a sampling module when the instrument is powered-on.
Always use a wrist strap (provided with your instrument) when handling
sampling modules or making signal connections. Wear anti-static clothing and
work in a static-free workstation when using sampling modules.