38 Chapter 3
Test Set Cal and SelfCal
Test Set Cal: An Overview
Step 3. Press Test Set Cal.
Step 4. Press
Calibration.
NOTE If Calibration is disabled, the status of the multi-port test set of this channel may be
off. See "Procedure of setting ports for the multi-port test set" in the previous
chapter.
Step 5. Press
Port x-y (a port pair for which you want to execute calibration).
NOTE The multi-port test set calibration is performed for each port pair. If you want to use,
for example, ports 1 and 3 for measurement, calibration must be performed for 1-2
and 3-4.
Step 6. Connect the open standard to the Port x side and press
Port x Open to measure
calibration data.
Step 7. Connect the short standard to the Port x side and press
Port x Short to measure
calibration data.
Step 8. Connect the load standard to the Port x side and press
Port x Load to measure
calibration data.
Step 9. Connect the open standard to the Port y side and press
Port y Open to measure
calibration data.