70 Chapter 6
SCPI Command Reference
:SENS{1-4}:CORR:MULT:COLL:ECAL
:SENS{1-4}:CORR:MULT:COLL:ECAL
Syntax :SENSe{[1]|2|3|4}:CORRection:MULTiport:COLLect:ECAL <value 1>,<value 2>
Description For channel 1 to channel 4, performs the calibration of the specified ports using
ECal. This command makes a measurement of the standard and calculates
calibration coefficient successively. The calibration executed here is the test set
calibration specified by :SENS{1-4}:CORR:MULT:COLL:METH on page 73.
NOTE If 2 ports you specify are not a port pair, an error occurs. Also, if
:SENS{1-4}:MULT on page 81 is off, an error occurs.
Parameters
Query response No query
Example of use
10 OUTPUT 717;":SENS1:CORR:MULT:COLL:ECAL 1,2"
Related commands :SENS{1-4}:MULT on page 81
:SENS{1-4}:CORR:MULT:COLL:METH on page 73
Equivalent key
[Cal] - Test Set Cal - Calibration - Port x-y - Ecal
Description
value 1 Specifies a port of the test set port pair to be calibrated.
value 2 Specifies the other port of the test set port pair to be
calibrated.