Agilent Technologies 87075C Network Router User Manual


 
Chapter 6 73
SCPI Command Reference
:SENS{1-4}:CORR:MULT:COLL:METH
:SENS{1-4}:CORR:MULT:COLL:METH
Syntax :SENSe{[1]|2|3|4}:CORRection:MULTiport:COLLect:METHod {ERES|SOLT2}
:SENSe{[1]|2|3|4}:CORRection:MULTiport:COLLect:METHod?
Description For channel 1 to channel 4, selects the test set calibration from enhanced response
calibration or full 2-port calibration.
NOTE Only ERES (enhanced response calibration) can be specified for the T/R model.
Parameters
Example of use
10 OUTPUT 717;":SENS1:CORR:MULT:COLL:METH SOLT2"
20 OUTPUT 717;":SENS1:CORR:MULT:COLL:METH?"
30 ENTER 717;A$
Related commands :SENS{1-4}:CORR:MULT on page 69
Equivalent key
[Cal] - Test Set Cal - Calibration - Cal Type
Description
ERES Specifies the enhanced response calibration for the test set
calibration.
SOLT2 Specifies the full 2-port calibration for the test set
calibration.