Agilent Technologies 87075C Network Router User Manual


 
80 Chapter 6
SCPI Command Reference
:SENS{1-4}:CORR:MULT:SELF:TIM
:SENS{1-4}:CORR:MULT:SELF:TIM
Syntax :SENSe{[1]|2|3|4}:CORRection:MULTiport:SELF:TIMer <value>
:SENSe{[1]|2|3|4}:CORRection:MULTiport:SELF:TIMer?
Description For channel 1 to channel 4, specifies the interval at which the self calibration of the
test set is executed automatically. If :SENS{1-4}:CORR:MULT:SELF on page 78
is on, self calibration is executed at intervals specified by this command.
Parameters
Query response <value><newline><^END>
Example of use
10 OUTPUT 717;":SENS1:CORR:MULT:SELF:TIM 30"
20 OUTPUT 717;":SENS1:CORR:MULT:SELF:TIM?"
30 ENTER 717;A
Related commands :SENS{1-4}:MULT on page 81
:SENS{1-4}:CORR:MULT:SELF on page 78
Equivalent key
[Cal] - Test Set Cal - Timer
<value>
Description Interval for auto run of self calibration.
Range 1 to 43200
Preset value 60
Unit Minute