Agilent Technologies G1978B Computer Accessories User Manual


 
16 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide
1 Installation
To convert from ESI, APCI or APPI to the multimode source
Figure 11 Standard spray shield and capillary cap for ESI or APCI
10 Remove the capillary cap. If needed, moisten a clean cloth with isopropyl
alcohol and wipe the capillary cap. See Figure 12.
Figure 12 Spray shield removed.
11 Place the capillary cap back on the capillary.
12 Install the new spray shield with field shaping electrodes. See Figure 13.
Standard spray shield
Capillary cap
Capillary cap