Agilent Technologies G1978B Computer Accessories User Manual


 
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide 31
Verification 2
To verify sensitivity for the G1978B Multimode using Multiple FIA method in Mixed Mode operation
d Click OK.
8 Repeat for the MMCheckSL_CI.M data file. Use the same subdirectory and
change Data File to Multi_CI.
9 Repeat for the MMCheckSL_MX.M data file. Use the same subdirectory and
change Data File to Multi_MX as the data file name.
10 Click Run to start the sequence.
11 Review the results.
When the last method has finished, the report “Multimode Verification Report”
prints.