20 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide
1 Installation
To convert from ESI, APCI or APPI to the multimode source
18 Connect the LC/MS sample tubing to the LC/MS diverter valve inlet filter.
See Figure 19 on page 20.
Figure 19 LC/MS sample tubing connected to LC/MS inlet filter
19 If you are installing the multimode source for the first time, follow the steps
in “Step 2. Install the HV control PCA and cables” on page 9.
WARNING
The LC/MS Liquid Chromatograph diverter valve is an integral part of the G1978B
safety system. The LC mobile phase flow must always be connected to the diverter
valve inlet filter. Never bypass the diverter valve and connect directly to the
nebulizer. If the diverter valve is used in a manner not specified by Agilent
Technologies, the protections provided by the diverter valve may be impaired.
LC/MS sample tubing