Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide 29
Verification 2
To verify sensitivity for 6130 and 6140 Quads - G1978B Multimode, Positive Scan Modes
When the method is finished, a report prints showing the signal-to-noise
ratio for the five blank and five sample peaks, and a blank-subtracted
average of the sample peaks. The blank-subtracted average signal-to-noise
value must be greater than or equal to 20:1 peak-to-peak (100:1 RMS).
Sensitivity Verification for Multimode in APCI, Positive Scan Mode
1 Start the ChemStation software.
2 Change the view to MSD Tune, and start an Autotune.
After the autotune completes, you may need to wait up to 30 minutes to
allow for the calibrant solution to be pumped out of the Quad instrument.
This minimizes any background signal resulting from the calibrant.
3 Change the view to Method and Run Control, and load the method
6130SCNCI_MM.M (6130 model) or 6140SCNCI_MM.M (6140 model).
4 Edit the method to ensure that 75:25 methanol/water with 5mM ammonium
formate is selected as the LC solvent.
5 Place the vials into the LC autosampler.
Position #1: empty, uncapped vial
Position #2: vial of the solvent used for dilution (solvent blank)
Position #3: vial with the reserpine sample (10 pg/µL)
6 Run the method.
The method does an FIA run with one injection of the empty vial, five
injections of the solvent blank, and five injections of the reserpine sample.
You may need to further optimize the nebulizer pressure by running FIA
experiments to achieve maximum instrument sensitivity
7 Review the results.
When the method is finished, a report prints showing the signal-to-noise ratio
for the five blank and five sample peaks, and a blank-subtracted average of the
sample peaks. The blank-subtracted average signal-to-noise value must be
greater than or equal to 10:1 peak-to-peak (50:1 RMS).