Fluke 2625A Power Supply User Manual


 
2620A, 2625A
Users Manual
4-44
Table 4-8. Command and Query Reference (cont)
Use the lowest two bits of the value given as individual flags to specify
the temperature configuration. Therefore, the value given must be in the
range from 0 through 3 or an Execution Error is generated. These
settings affect every channel; they cannot be set for each channel
individually. If this command is attempted when measurements are
active, an Execution Error results.
Bit Number Value Meaning
00°C
1 °F
1 0 Disable open thermocouple detection
1 Enable open thermocouple detection
TEMP_CONFIG? Temperature Configuration Query
Response to this query corresponds to settings made with the
TEMP_CONFIG command, as follows:
Bit Number Value Meaning
00°C
1 °F
1 0 Disable open thermocouple detection
1 Enable open thermocouple detection
TIME Set the instrument clock.
TIME <hours>,<minutes>
<hours> = (0 .. 23) (24-hour scale, 18:00 = 6:00 pm).
<minutes> = (0 .. 59)
Invalid values generate an Execution Error. Seconds are automatically
set to 00 with the TIME command.
TIME_DATE? Retrieve Time and Date
Returns comma-separated integer values for time, date, and year as
maintained by the instrument clock. These returned values use the TIME
and DATE command integer formats as follows:
hours 0-23
minutes 0-59
seconds 0-59
month 1-12
day 1-31
year 00-99
Seconds are returned, but cannot be set other than to 00.