Fluke 2625A Power Supply User Manual


 
2620A, 2625A
Users Manual
6-4
Power-Line Cord Connector
Line Fuse
(T 125 mA, 250V,
Slow Blow)
Fuse Holder
(Spare Fuse Provided)
Complies with the limit for a class B computing device
pursuant to Subpart J of Part 15 of FCC Rules
+–0123TR
0123
4567
Σ
9-16 V
DC PWR
!
+30V
DIGITAL I/O
MEETS 0871 B
To Remove,
Squeeze and
Slide Out
oo18f.eps
Figure 6-1. Replacing the Line Fuse
Performance Tests
When received, the 2620A/2625A Hydra Series II instrument is calibrated and in
operating condition. The following Performance Verification Procedures are provided
for acceptance testing upon initial receipt or to verify correct operation at any time. All
tests may be performed in sequence to verify overall operation, or the tests may be run
independently.
If the instrument fails any of these performance tests, calibration adjustment and/or
repair is needed. To perform these tests, you will need a Fluke 5700A Multifunction
Calibrator or equipment meeting the minimum specifications given in Table 6-2.