Fluke 2645A Network Card User Manual


 
Maintenance
Performance Test
6
6-7
Table 6-2. Recommended Test Equipment
Instrument Type Minimum Specifications Recommended Model
Multifunction DC Voltage: Fluke 5700A
Calibrator
Range: 90 mV to 300V dc
Accuracy: 0.002%
AC Voltage:
Frequency Voltage Accuracy
1 kHz 29 mV to 300V 0.05%
100 kHz 15 mV to 300V 0.5%
Frequency:
10 kHz 1V rms 0.01%
Ohms:
Ohms Accuracy
190 0.005%
1.9 k 0.005%
19 k 0.005%
190 k 0.005%
1.9 M 0.005%
Mercury Thermometer 0.02°C Resolution Princo ASTM-56C
Thermocouple Probe Type T Fluke P20T
Oil/Water Bath Thermos bottle and cap
Digital Multimeter General Purpose Measurement Fluke 77
Signal Generator Sine wave. 0.5 to 1V rms, 10 Hz to 5 kHz Philips PM5193
Alternative Equipment List
Instrument Type Recommended Model
DC Voltage Calibrator Fluke 5440B
DMM Calibrator Fluke 5100B (AC Volts Only)
Function/Signal Generator Philips PM5139 or HP33120A
Decade Resistance Source General Resistance RDS 66A