Service Calibrations 4-29
7. Turn the sensor clockwise one-half turn to achieve a smaller
range reading, or counterclockwise to achieve a larger sen-
sor range reading.
8. Tighten the two nuts that secure the media thickness
sensor.
9. Press ! (Proceed) to run the sensor’s height test again (see
step 2 on page 4-28 for instructions).
10. Repeat the adjustment of the sensor’s height until the sen-
sor is within optimal range.
11. Press ! (Proceed) to calibrate the service station as shown
in “Calibrating the Service Station” on page 4-30.