HP (Hewlett-Packard) TM 11-6625-2779-14&P Computer Accessories User Manual


 
PERFORMANCE CHECK TEST CARD
Hewlett-Packard Model 3555B Tests Performed By
Transmission and Noise Measuring Set
Serial No. Date
DESCRIPTION CHECK
CARRIER 75 UNBAL
LEVEL ACCURACY CHECK
30Hz to 1MHz
+10dBm Range +10dBm ±0.5dBm
0dBm Range 0dBm ±0.5dBm
-10dBm Range -10dBm ±0.5dBm
-20dBm Range -20dBm ±0.5dBm
-30dBm Range -30dBm ±0.5dBm
-40dBm Range -40dBm ±0.5dBm
-50dBm Range -50dBm ±0.5dBm
100Hz to 600kHz
+10dBm Range +10dBm ±0.2dBm
0dBm Range 0dBm ±0.2dBm
-10dBm Range -10dBm ±0.2dBm
-20dBm Range -20dBm ±0.2dBm
-30dBm Range -30dBm ±0.2dBm
-40dBm Range -40dBm ±0.2dBm
-50dBm Range -50dBm ±0.2dBm
1MHz to 3MHz
+10dBm Range +10dBm ±0.5dBm ±10% of meter indication
0dBm Range 0dBm ±0.5dBm ±10% of meter indication
-10dBm Range -10dBm ±0.5dBm ±10% of meter indication
-20dBm Range -20dBm ±0.5dBm ±10% of meter indication
-30dBm Range -30dBm ±0.5dBm ±10% of meter indication
40dBm Range -40dBm ±0.5dBm ±10% of meter indication
-50dBm Range -50dBm ±0.5dBm ±10% of meter indication
CARRIER 135 BAL
LEVEL ACCURACY CHECK
1kHz to 600kHz
+10dBm Range +10dBm ±0.5dBm
0dBm Range 0dBm ±0.5dBm
-10dBm Range -10dBm ±0.5dBm
-20dBm Range -20dBm ±0.5dBm
-30dBm Range -30dBm ±0.5dBm
-40dBm Range -40dBm ±0.5dBm
-50dBm Range -50dBm ±0.5dBm
10kHz to 300kHz
+10dBm Range +10dBm ±0.2dBm
0dBm Range 0dBm ±0.2dBm
-10dBm Range -10dBm ±0.2dBm
-20dBm Range -20dBm ±0.2dBm
-30dBm Range -30dBm ±0.2dBm
-40dBm Range -40dBm ±0.2dBm
-50dBm Range -50dBm ±0.2dBm
(1)