53
Bar Code Calibration Procedure without Remote Display
11. Remove half of the load and center the remaining load. The blue LED will blink four times periodically.
Wait for scale stability, and then press the beeper volume switch once. The beeper will beep 4 times
indicating that the Cal 4 value has been stored.
Figure 56.
12. Remove the remaining half load from the scale. The blue LED will blink five times periodically. Wait for
scale stability, and then press the beeper volume switch once. The beeper will beep 5 times indicating that
the Cal 5 value has been stored.
Figure 57.
13. Calibration is now complete. The scanner will automatically restart, and beep one time, in 5 seconds.
14. In order to use the scale in the normal operating mode, the unit must be powered down and the scale
program switch returned to the run position.
Figure 58.
See Power Source caution statement located on page 12 of this manual.
15. The unit’s calibration must now be verified as required by state and/or local Weights and Measures
regulations (starting on page 54).