National Instruments NI 783xR Switch User Manual


 
Appendix A Specifications
NI 783xR User Manual A-8 ni.com
Physical
Dimensions (not including connectors)
NI PXI-783xR..................................16 cm × 10 cm (6.3 in. × 3.9 in.)
NI PCI-783xR..................................17 cm × 11 cm (6.7 in. × 4.3 in.)
I/O connectors.........................................Three 68-pin female high-density
VHDCI type
Maximum Working Voltage
Maximum working voltage refers to the signal voltage plus the
common-mode voltage.
Channel-to-earth .....................................±12 V, Measurement Category I
Channel-to-channel.................................±24 V, Measurement Category I
Caution Do not use the NI 783xR for connection to signals in Measurement Categories II,
III, or IV.
Environmental
The NI 783xR is intended for indoor use only.
Operating Environment
NI 7830R, NI 7831R
40 MHz or 80 MHz timebase..........0 °C to 55 °C, tested in
accordance with IEC-60068-2-1
and IEC-60068-2-2
NI 7833R
40 MHz timebase.............................0 °C to 55 °C, tested in
accordance with IEC-60068-2-1
and IEC-60068-2-2
80 MHz timebase.............................0 °C to 55 °C except the
following: 0 °C to 45 °C when
installed in an NI PXI-1000/B or
NI PXI-101X; tested in
accordance with IEC-60068-2-1
and IEC-60068-2-2