Unpacking and Setting Up the Model 651 Teledyne API Ultrafine Particle Monitor - Model 651
26 07506C DCN6727
To connect the water supply, follow these instructions:
1. Using a 7/64 inch hex driver, mount the water supply bottle
bracket to the front or back of the particle counter using the
provided bottle bracket mounting screws. The figure below
shows the bracket mounted on the back.
Figure 2-1
Connecting the Water Supply
2. Fill the water supply bottle with either distilled (<6 ppm) or
HPLC water and place the bottle in the bracket.
Note: A filled water supply bottle will typically allow the Model
651 to operate for more than the 4-weeks wick
replacement interval. If water is added between the wick
change, it is recommended that the water be added to the
bottle without disconnecting it from the Model 651 to avoid
adding any bubbles into the water supply line.
3. Push the connector on the water supply bottle tubing into the
WATER FILL fitting on either the front or back of the
instrument (figure below shows the back).
Bracket
Screw