Xantrex Technology XDC 20-300 Power Supply User Manual


 
Entering Calibration Mode
TM-XDOP-01XN B–3
Entering Calibration Mode
Calibration mode can be entered from the front panel by selecting
"CALIBRATION" from the main menu or by using the "change calibration state"
SCPI command.
Front Panel
To access calibration mode via the front panel:
If your password code is incorrect, the prompt displays Incorrect code and
automatically returns to Code ####. Either try again or press EXIT to escape.
If your password code is correct, the prompt displays Output V Cal. This is
the first of 12 available options. The 12 sets of parameters that must be adjusted
during calibration are:
Output V Cal: Calibrate voltage output and readback.
Output I Cal: Calibrate current output and readback.
ANLG V PGM 5V: Calibrate 5V analog programming interface for setting
voltage output.
ANLG V RB 5V: Calibrate 5V analog programming interface for monitoring
voltage output.
ANLG I PGM 5V: Calibrate 5V analog programming interface for setting
current output.
ANLG I RB 5V: Calibrate 5V analog programming interface for monitoring
current output.
CAUTION
Calibration procedures should only be performed by qualified users. Failure to adhere to
this warning may cause damage to the power supply, or pose a safety hazard for the user.
Step # Do This You Will See
1ERROR MSGS
2 Press 2 times. CALIBRATION
3 Code #####
4 Enter the calibration security code. The
factory code is “0000”.
Code 0000
5
You are now in the
Calibration menu.
Output V Cal
MENU
ENTER
ENTER
TM-XDOP-01XN.book Page 3 Monday, July 17, 2006 11:19 AM