Agilent Technologies 1670G Network Hardware User Manual


 
123
Triggering Examples
Cross-Arming Trigger Examples
To capture the waveform of a glitch using the
oscilloscope (oscilloscope option only)
The following setup uses the triggering capability of the timing
analyzer and the acquisition capability of the oscilloscope.
1 Set up a timing analyzer. Go to the timing analyzers Format
menu and set the Timing Acquisition Mode to Half Channel
500 MHz.
2 Go to the timing analyzers Trigger menu.
3 Select an Edge term. Then assign glitch detection * to the
channels of interest represented by the Edge term.
4 Go to the Arming Control menu. Set the Scope Arm In to
Analyzer.
5 Select Group Run in the Analyzer Arm In menu.
6 Go to the Scope Trigger menu, and set Mode to Immediate.
If you have trouble capturing the glitch waveform on the oscilloscope,
try adjusting the skew in the Arming Control menu, so the oscilloscope
triggers earlier.
NOTE: A timing analyzer can trigger on a glitch and capture it, but a timing analyzer
doesnt have the voltage or timing resolution to display the glitch in detail. An
oscilloscope can display a glitch waveform with fine resolution, but cannot
trigger on glitches, combinations of glitches, or sophisticated patterns
involving many channels.