![](http://pdfasset.owneriq.net/2/8e/28e11347-f9d5-476f-8c53-845e906a3d11/28e11347-f9d5-476f-8c53-845e906a3d11-bg53.png)
83
Using the Trigger Menu
Setting Up Time Correlation between Analyzers
To set up time correlation between a timing
and a state analyzer
To set up time correlation between a timing and a state analyzer, only
the state analyzer needs to have Count Time turned on. The timing
analyzer automatically keeps track of time.
1 In the state Analyzer Trigger menu, select Count.
Count may be Count Off, Count Time, or Count States. Selecting the
field causes a pop-up to appear.
2 Select the field after Count: and select Time.
A warning may appear about reduced memory. It will not prevent you
from changing Count to Count Time.
3 Select Done.
Now when you acquire data you will be able to set up a mixed display.