504
Using the LAN’s X Window Interface
Using the LAN’s X Window Interface
To close the interface
From the XWindow Interface or Front Panel
1 Go to the System External I/O menu.
2 Select the Disconnect field.
The interface on your Xserver closes, and the Disconnect field changes
to Connect.
From a Remote Computer
• Send the XWIN OFF programming command to the logic analyzer.
NOTE: Note that simply closing the window that the interface is running in may leave
the logic analyzer hung up.
Example
ftp method
To end the session using ftp, create a second text file with the following
contents:
xwin off<cr>
The second text file is named "stopx". At an ftp command prompt, copy
this file from your local directory to a virtual file called "program" in the
logic analyzer \system directory as follows:
ftp> cd system
ftp> put stopx program