Agilent Technologies 6010a Power Supply User Manual


 
26
load resistor that is capable of safely dissipating 1000 watts. Proceed as follows:
a. Connect the test equipment as shown in Figure 2-8.
b. Switch the unit's power-on and turn the output voltage all the way up.
c. Turn up output current to:
17.0Adc (6010A)
120Adc (6011A)
50Adc (6012B)
5.0Adc (6015A)
Check that the unit's CC LED remains lighted.
d. Check that the rms noise current measured by the current probe and rms voltmeter is no more than:
15mA rms (6010A).
120mA rms (6011A)
25mA rms (6012B)
50mA rms (6015A)
Figure 2-8. CC PARD Test Setup