Agilent Technologies E1439 Network Card User Manual


 
175
Agilent E1439 Programmer's Reference
Functions listed alphabetically
age1439_state_save
Saves the module’s current instrument state.
age1439_state_save
VXIplug&play Syntax
#include "age1439".h
ViStatus age1439_state_save(ViSession id);
Description
This function may be used to save a state to which you want to return later. age1439_reset does
not change a saved state. The state is not saved to non-volatile RAM.
Note The saved state is lost by issuing the following commands: age1439_input_range_auto, age1439_
input_autozero, age1439_self_test,andage1439_reset_hard.
Parameters
id is the VXI instrument session pointer returned by the age1439_init function.
Return Value
AGE1439_SUCCESS indicates that a function was successful.
Values other than AGE1439_SUCCESS indicate an error condition or other important status
condition. To determine the error message, pass the return value to “age1439_error_message” on
page 102.
See Also