Barco 6MP Projector Accessories User Manual


 
32 Coronis Fusion 4MP/6MP
ALC & DICOM OPTIONS SUBMENU
Measured ambient
light
Indicates the ambient light actually
measured by the ambient light sensor on
the front of the display. This is a read-only
value.
Averaged ambient
light
Shows the average of the measured
ambient light since the display was
switched on.
When you execute the function “Measure
ALC Value”, this value will be stored as the
Measured ALC correction value for the
selected reading room, unless this value is
higher than the maximum ambient light
determined for the selected reading room.
Correction value Shows the ambient light correction value
that is taken into account in the calculation
of the display function.
Cont. DICOM ALC Allows to switch continuous DICOM ALC
on/off.
When switched on, the DICOM DF is
recalculated continuously taking the
averaged ambient light into account.
When switched off, the DICOM DF is
recalculated only at the moment the
DICOM Offset is changed (see below).