EMC notice
No specific requirement on the use of external cables or other
accessories except power supply.
With the installation of the device, use only the delivered power supply
or a spare part provided by the legal manufacturer. Using another can
result in a decrease of the immunity level of the device.
ELECTROMAGNETIC EMISSIONS
The CORONIS FUSION 4MP/6MP display is intended for use in the
electromagnetic environment specified below.
The customer or the user of the C
ORONIS FUSION 4MP/6MP display should
assure that it is used in such an environment.
This C
ORONIS FUSION 4MP/6MP display complies with appropriate medical
EMC standards on emissions to, and interference from surrounding
equipment. Operation is subject to the following two conditions: (1) this
device may not cause harmful interference, and (2) this device must
Emissions test Compliance Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1 The C
ORONIS FUSION 4MP/6MP display uses
RF energy only for its internal function.
Therefore, its RF emissions are very low and
are not likely to cause any interference in
nearby electronic equipment
RF emissions
CISPR 11
Class B
The CORONIS FUSION 4MP/6MP display is
suitable for use in all establishments,
including domestic establishments and
those directly connected to the public low-
voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class D
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies