Barco 6MP Projector Accessories User Manual


 
Coronis Fusion 4MP/6MP 47
Reading room definition
This function is available only when the DICOM display function is
selected and Continuous DICOM ALC is off.
TO MODIFY A READING ROOM DEFINITION:
1 Select the DICOM display function. See To select a display
function (DF):, on page 43.
2 Switch Continuous DICOM ALC off. See To switch continuous
DICOM ALC on/off:, on page 44.
3 In the ALC & DICOM Options menu, select Reading Room Def.
4 Select the reading room definition you wish to modify. Thereto,
first select Reading Room. Select the appropriate reading room.
5 To modify the maximum ambient light for this reading room,
select Max. Ambient Light and change the value.
You can change this value within certain limits determined by the
selected reading room.
6 To modify the Preset correction value for this reading room, select
Preset Corr. Value and change the value.
You can change this value within certain limits determined by the
selected reading room.
7 To modify the Measured ALC Value for this reading room, select
Measure ALC Value and execute the function.
If the “Averaged Ambient Light” at that moment is lower than the
defined “Maximum Ambient Light”, the averaged ambient light
value is copied as Measured ALC value.
If the “Averaged Ambient Light” at that moment is higher than the
defined “Maximum Ambient Light”, the display will prompt a
warning for the user to take measures to adapt the ambient light
in the room to the selected reading room. Proceed as follows:
Use Enter or click the control wheel. As a result the
Measured ALC value will be limited to the maximum ambient