Barco 6MP Projector Accessories User Manual


 
Coronis Fusion 4MP/6MP 43
Display function selection
TO SELECT A DISPLAY FUNCTION (DF):
1 Open the main menu.
2 Open the Display Function menu.
3 Select Display Function.
4 Select the desired DF.
5 Return to the Display Function menu.
You can select the following DFs:
DICOM Select a DICOM display function for most
medical viewing applications.
The DICOM function results in more visible
grayscales in the images.
Dynamic Gamma 2.2
Dynamic Gamma 1.8
These are gamma functions that are shifted
to take into account the non-zero
luminance of an LCD panel when driven
with a “black” signal.
They are especially useful in CT applications
to improve the perception of low
Hounsfield values.
Native If you select Native, the native panel
behavior will not be corrected.
Test This is identical to the Native DF.
Gamma 2.2
Gamma 1.8
Select one of these display functions in
case the display is to replace a CRT display
with a gamma of 1.8 or 2.2 respectively.
User This is a user-definable DF.