Wedge Quiet Interval
Product Reference Guide 85
Wedge Quiet Interval
This option specifies the amount of time to look for keyboard activity before the reader breaks
the keyboard connection in order to transmit data to host. The selectable range for this feature is
from 0 to 990ms in 10ms increments.
NOTE
Follow these instructions to set this feature:
1. Determine the desired setting in milliseconds.
2. Divide the desired setting by 10 (setting is in 10ms increments). Pad the result with lead-
ing zeroes to yield two digits. For example: 0 = 00, 5 = 05, 20 = 20, etc.
3. Scan the ENTER/EXIT PROGRAMMING MODE barcode to enter Programming
Mod
e.
4. Scan the barcode: SELECT WEDGE QUIET INTERVAL SETTING.
5. Scan the appropriate two digits from
the keypad in Appendix E
, Keypad, that represent the
duration which was determined in the steps abov
e. You will hear a two-beep indication
after the last character.
NOTE
6. Scan the ENTER/EXIT PROGRAMMING MODE barcode to exit Programming
Mod
e.
This completes the procedure to set the Wedge Quiet Interval. See Table 14 for some examples
of how to set this feature.
Table 14. Timeout Setting Examples
This feature applies ONLY to the Keyboard Wedge interface.
If you make a mistake before the last character, scan the CANCEL
bar
code to abort and not save the entry string. You can then start
again at the beginning.
STEP ACTION EXAMPLES
1 Desired Setting
10ms 150ms 600ms 850ms
2
Divide by 10 (and pad with
leading ze
roes)
01 15 60 85
3 Scan ENTER/EXIT PROGRAMMING MODE
4 Scan SELECT WEDGE QUIET INTERVAL SETTING
5
Scan Two Characters From
Appendix E, Keypad
‘0’ and ‘1’ ‘1’ and ‘5’ ‘6’ and ‘0’ ‘8’ and ‘5’