Emerson 5081-T Switch User Manual


 
54
MODEL 5081-T SECTION 7.0
PROGRAMMING
1. Press PROG on the remote controller.
2. Press NEXT until the diSPLAy submenu appears. Press ENTER.
3. Press or
to display the desired measurement. Press ENTER to save.
4. tEMP will appear. Press or
to display the desired temperature reading, C or F.
Press ENTER.
5. OUtPUt will appear. Press or
to display the desired 4-20 output value, Cur or %.
Press ENTER.
6. COdE will appear. Using the arrow keys on the remote control, enter the desired 3-digit
security code for accessing the Calibration, Program, and Diagnostic functions via the
Remote Control handheld. Press ENTER. The dISPLAY mnemonic will appear.
PROGRAM
UA:1
EXIT ENTER
/QPGVF
PROGRAM
G5<:[-A
EXIT NEXT ENTER
7.4 DISPLAY
7.4.1 Purpose
This section describes how to do the following:
1. Configure the transmitter to measure conductivity, resistivity, or set up a custom curve
2. Set the temperature units to °C or °F
3. Set the output to current or percent of full scale
4. Enter a security code.
7.4.2 Definitions
1. MEASUREMENT. The transmitter can be configured to measure conductivity in mS/cm or resistivity in MegOhms, or
configured with a 3-5 point custom curve for special applications.
2. OUTPUT CURRENT. The transmitter generates a 4-20 mA output signal directly proportional to the conductivity or resis-
tivity of the sample. The output signal can be displayed as current (in mA) or as percent of full scale.
5. SECURITY CODE. The security code unlocks the transmitter and allows access to all menus.
7.4.3 Procedure
COnduc Conductivity
nAOH Sodium Hydroxide 0-15%
HCL Hydrochloric Acid 0-16%
H2SO4L Sulfuric Acid 0-30%
H2SO4H Sulfuric Acid 96-99.7%
CuSt Custom Curve