MODEL 5081-T SECTION 7.0
PROGRAMMING
7.7 RANGE
7.7.1 Purpose
This section provides the steps to select automatic ranging or a specific fixed range of measurement. Five specific con-
ductance ranges are selectable. Setting the Model 5081 to a fixed range reduces response time.
The following conductance ranges are available:
NOTE: The selection between automatic ranging or a specified fixed range of measurement range can ONLY be done
using the IRC/Infrared Remote Controller. This selection cannot be done via HART or F
OUNDATION Fieldbus hosts or
configurators.
7.7.2 Procedure
1. Press PROG.
2. Press NEXT five times.
S-PJU!
!
(range) will appear.
3. Press ENTER. The default setting Auto will appear. This indicates that Model 5081 is in auto ranging mode.
4. To set a fixed conductance range, press the down arrow key
until the desired measurement range appears.
1400mS, 550mS, 200mS, 33mS and 3mS will appear when pressing the down arrow successively.
5. When the desired range is reached, press ENTER. This disables auto ranging and limits the measurement to the
selected range.
6. Press NEXT to move to the next Programming menu item.
Setting Measurement Range Over Range Warning Limit
Auto 0uS to 1400mS None
1400mS 550mS to 1400mS None
550mS 200mS to 550mS 570mS
200mS 33mS to 200mS 207mS
33mS 3000uS to 33mS 34mS
3mS 0uS to 3000uS 3400uS
1. Press PROG on the remote controller.
2. Press NEXT until the dEFAULt appears in the display. Press ENTER.
3. Use or
to toggle between nO and yES. With yES showing, press ENTER to return
to factory default settings.
PROGRAM
2-FUQSA
EXIT ENTER
P9
PROGRAM
G12->[U
EXIT NEXT ENTER
7.8 DEFAULT
7.8.1 Purpose
This section describes how to erase ALL user-defined configuration settings and return the transmitter to factory default
settings. All custom curve values and settings will be deleted.
7.8.2 Procedure
57