Fluke 9933 Calculator User Manual


 
If the third line does not exist, this software does not compensate the voltage
readings.
Example: ASCII text file for Thermocouple coefficients.
–100.0,0.0,100.0,200.0,300.0
–5506.65,0.02,6322.54,13413.47
–0.02,–0.02,0.01,–0.02,–0.02
Note: If a specific set of temperature set-points are used repeatedly, an
import/export file can be created which contains only the temperatures.
The line(s) that normally contain the resistance or voltage information can
be left blank. From the Raw Data dialog, select the Import button to im
-
port the file, then enter the resistance data.
3.5 Coefficients and residuals
The Coefficients and Residuals dialog (Figure 19) is displayed when the Calcu-
late button is selected from the Raw Data dialog.
Enabled and disabled
The Set-points and Residuals list is disabled when calculating IPTS-68 coeffi
-
cients because residuals are not calculated.
29
3 Coefficients menu
Coefficients and residuals
Figure 19 Coefficients and Residuals dialog