Fluke DSP-100 Network Card User Manual


 
DSP-100,DSP-2000 Users Manual Supplement
1/99 1
Change #1
On page 6-1, under
Calibrating the Test Tool,
add the following note after
the first paragraph:
Note
Allow a 1 minute warm-up time prior to performing self-calibration (see
Chapter 6). Self-calibration should be performed only after the test tool
and remote unit have warmed up to an ambient temperature between
10°C and 40°C (50
°
F and 104
°
F).
On page 8-11, replace the
Return Loss
specification with:
Return Loss (RL)
Return Loss range: 0 to 30 dB
Accuracy: ± 3 dB over the following ranges:
0 to 18 dB from 5 to 10 MHz
0 to 15 dB from 10.1 to 20 MHz
0 to 10 dB from 20.1 to 100 MHz